Fully Auto Biochemistry Analyzer, BA-A-200TC

  • Throughput: 200 tests/hour
  • Sample Volume: 1.6–50 µL, 0.1 µL step
  • 40 sample positions, 40 reagent positions and 60 reusable plastic cuvettes
  • Wavelength: 340–690 nm, 9 wavelengths, accuracy ±1.5 nm

Share the product

Catégorie :

Description

Sample & Reagent Positions

  • 40 sample positions
  • Reagent refrigeration function
  • 40 reagent positions

Reaction Cuvettes

  • 60 reusable plastic cuvettes
  • Automatic cuvette blank testing
  • Rear spectrophotometry optical system
  • Detachable design for easy replacement

Sample/Reagent Probe & Mixer

  • Automatic liquid level detection
  • Collision protection function
  • Interior and exterior probe washing
  • Independent mixer

7-Step Auto-Washing System

  • Automated 7-step washing
  • Advanced wash station ensures accurate results
  • Low water consumption: 1–2 L/h

Intelligent Software

  • Automatic cuvette washing at startup and shutdown
  • Automatic reagent volume detection at startup
  • Reagent tray temperature detection at startup

Cuvette Blank & Calibration

  • Automatic cuvette blank measurement
  • Automatic washing of selected cuvettes

Real-Time Operating Status Display

  • Real-time monitoring of sample tray, reagent tray, and cuvettes
  • Real-time tracking of residual reagent volumes
  • Real-time monitoring of liquid waste volume

Automatic Dilution and Retest

  • Automatic dilution and retest when absorbance is out of range
  • Automatic dilution and retest when linearity limit is exceeded
  • Flexible setting for auto-dilution ratio and dilution water position

Specifications

Model BA-A-200TC
System Functions
Analysis Methods End-point, kinetic, two-point, double-reagent, double-wavelength, multi-standard, etc.; open to various reagents
Optimal Replacement for Semi-Auto Chemistry Analyzers
Minimal labor required;
Independent mixer;
Reduced reagent consumption: 300 µL/test
Throughput: 200 tests/hour
User-friendly: Smart and easy-to-operate software
High accuracy: Minimized operational and random errors
Proven reputation among clinics
Sample/Reagent Handling
Sample Volume 1.6–50 µL, 0.1 µL step
Sample Dilution Automatic/Manual pre-dilution, with dilution ratios up to 1:100
Reagent Positions 40
Reagent Volume 10–500 µL, 0.5 µL step
Reaction System
Reaction Cuvettes 60 reusable plastic cuvettes
Optical Length of Cuvette 6 mm
Reaction Volume 180–400 µL
Max. Reaction Time 10 minutes
Reaction Temperature 37 ± 0.1°C
Optical System
Light Source Halogen-tungsten lamp
Wavelength 340–690 nm, 9 wavelengths, accuracy ±1.5 nm
Absorbance Range 0–5.0 Abs
Resolution 0.0001 Abs
Optical Type Rear spectrophotometry
Calibration
Calibration Linear/non-linear multi-point calibration
Retest Automatically retests the sample when results exceed the linearity range or sample volume is insufficient
Operating System
Windows 10 or Windows 7
Input and Output
Input RS-232 interface / Computer
Output Multi-format printout
Working Conditions
Power Supply 100–240 V AC, 50/60 Hz, 300 VA
Ambient Temperature 10–35°C
Humidity ≤ 90%, non-condensing
Dimensions and Weight
Analyzer 548*413*360 mm, 27.3 kg
Package 910*680*675 mm, 39 kg